Shop online for a wide selection of Eppendorf 5424 Microcentrifuges Use these centrifuges for all modern molecular biology applications in Eppendorf™ and PCR tubes. Please contact customer service for assistance: 1-800-766-7000. 05-400-007 View Documents View Product Certificates. Centrifuge 5804/5804 R/5810/5810 R — Operating manual 1 Operating instructions 1 Operating instructions 1.1 Using this manual Please read this operating manual completely before using the device for the first time. Also observe the instructions for use enclosed with the accessories. Centrifuge 5424 / 5424 R — Operating manual 8 2 Product description Operating manual — Centrifuge 5424 / 5424 Rseeon p.Fig.Tab.p. 2.1 Main illustration The depiction of the Centrifuge 5424 / 5424 R can also be found on the front fold-out page (see Fig. 1: Front and rear view of the Centrifuge 5424.
Advertising
31
7 Troubleshooting
Centrifuge 5424 / 5424 R — Operating manual
7.2
Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies). 2. If necessary, repeat centrifugation.
Symptom / message
Cause
Remedy
LID ERROR
Centrifuge lid cannot be locked.
Try to close the centrifuge lid again.
LID ERROR
Centrifuge lid cannot be released.
1. Switch the device off and back on.
2. Press the open key. If the error occurs again:
1. Switch off the centrifuge.
2. Activate the emergency lid release
(see Emergency release on p. 32).
LID ERROR
Centrifuge lid must not be released during a
run. https://bossphones.weebly.com/blog/free-installer-download.
![]()
Wait for the rotor to stop.
LID LIFT
(only 5424 R)
Centrifuge lid has not been opened wide
enough.
Open the centrifuge lid wider by hand.
INT
Mains power failure during a run.
Check the mains connection.
NO RPM
Error in speed measuring system.
Leave the device switched on until the
error message disappears (10 s or 6 min).
Err 6
Drive error.
Repeat the run.
If this error message appears again,
switch centrifuge off and back on again after > 20 s.
Err 6
Drive overheated.
Allow the drive to cool down for at least 15
min.
Err 7
Major deviation in the speed control.
1. Wait for the rotor to stop.
2. Tighten the rotor.
Err 8
Drive error.
1. Wait for the rotor to stop.
2. Repeat the run.
Err 9 to 17
Electronics error.
Switch the centrifuge off and back on
again after > 20 s.
Err 18
(only 5424 R) ![]()
Too high temperature variation from nominal
value in the rotor chamber.
Check the settings.
Check unhindered air circulation through
the vents.
Thaw ice or switch off device and allow it
to cool down.
Err 19
(only 5424 R)
Cooling circuit is overheated.
Check unhindered air circulation through
the vents and allow device to cool down.
Err 20
(only 5424 R)
Temperature sensor in rotor chamber faulty.
Switch the centrifuge off and back on
again after > 20 s. Eppendorf Centrifuge 5424 Rotor Radius
Err 21
(only 5424 R)
Temperature sensor at capacitor faulty.
Switch the centrifuge off and back on
again after > 20 s.
Err 24
(only 5424 R)
Download drivermax for windows 8. Cooling unit fault, e.g., overheating.
Allow the centrifuge to cool down and
repeat the cycle.
EN
O
perating manual
Service Manual Bmw
Advertising
This manual is related to the following products:
Popular Brands
Popular manuals
5 Operation
Centrifuge 5424 / 5424 R — Operating manual
To change settings in the device menu, proceed as follows:
5.4.1Switch on centrifuge
Eppendorf 5415d Centrifuge Manual
1.Switch the centrifuge on, using the mains switch.
After switching on at the mains power switch, the centrifuge lid opens automatically.
2.Open the closed centrifuge lid by pressing the open key.
The parameter settings of the last run are displayed.
1.Place rotor vertically onto the motor shaft.
2.Insert the rotor key supplied into the rotor nut.
3.Turn rotor key clockwise until the rotor nut is firmly tightened.
1.Open the menu.
3.Confirm your selection.
4.Select the setting of the parameters in question.
Confirm the changed setting.
Eppendorf Centrifuge 5424 Service Manual
The BACK menu item belonging to the first menu level appears.
Hint!
To exit the second menu level without changing a parameter, select the BACK menu item and
CAUTION!
Eppendorf 5424 Service Manual Download
Load rotors symmetrically with identical tubes.
Always use tubes of the same type (weight, material/density and volume).
Check symmetric loading by balancing the adapters and tubes used with scales.
Operating manual